Abstract
A mediator-less microbial fuel cell (MFC) was used as a biochemical oxygen demand (BOD) sensor in an amperometric mode for real-time wastewater monitoring. At a hydraulic retention time of 1.05h, BOD values of up to 100mg/l were measured based on a linear relationship, while higher BOD values were measured using a lower feeding rate. About 60min was required to reach a new steady-state current after the MFCs had been fed with different strength artificial wastewaters (Aws). The current generated from the MFCs fed with AW with a BOD of 100mg/l was compared to determine the repeatability, and the difference was less than 10%. When the MFC was starved, the original current value was regained with a varying recovery time depending on the length of the starvation. During starvation, the MFC generated a background level current, probably due to an endogenous metabolism.
| Original language | English |
|---|---|
| Pages (from-to) | 607-613 |
| Number of pages | 7 |
| Journal | Biosensors and Bioelectronics |
| Volume | 19 |
| Issue number | 6 |
| DOIs | |
| State | Published - 15 Jan 2004 |
Keywords
- Biosensor
- BOD
- Continuous monitoring
- Microbial fuel cell
Fingerprint