Abstract
Water vapor is a pivotal obstacle when measuring ambient air pollutants. The effects of water vapor removal devices which are called KPASS (Key-compound PASSer) and Cooler. On the measurement of O3, SO2, and CO at ambient levels were investigated. Concentrations of O3, SO2, and CO were 100 ppb, 150 ppb, and 25 ppm, respectively. The amount of water vapor varied at different relative humidity levels of 30%, 50%, and 80% when the temperature was 25◦C and the pressure was 1 atm. Water vapor removal efficiencies and recovery rates of target gases were also determined. The KPASS showed a better performance than the Cooler device, removing 93.6% of water vapor and the Cooler removing 59.2%. In terms of recovery, the KPASS showed a better recovery of target gases than the Cooler. Consequently, it is suggested that the KPASS should be an alternative way to remove water vapor when measuring O3, SO2, and CO.
| Original language | English |
|---|---|
| Article number | 3446 |
| Journal | International Journal of Environmental Research and Public Health |
| Volume | 16 |
| Issue number | 18 |
| DOIs | |
| State | Published - Sep 2019 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Ambient air pollutants
- Carbon monoxide
- Cooler
- KPASS
- Ozone
- Sulfur dioxide
- Water removal device