Skip to main navigation Skip to search Skip to main content

Magnetic Resonance Electrical Impedance Tomography (MREIT): Phantom experiments for static resistivity images using J-substitution algorithm

  • Suk Hoon Oh
  • , Byung Il Lee
  • , Eung Je Woo
  • , Soo Yeol Lee
  • , Min Hyoung Cho
  • , Ohin Kwon
  • , Jeong Rock Yoon
  • , Jin Keun Seo
  • Yonsei University

Research output: Contribution to journal › Conference article › peer-review

4 Scopus citations

Abstract

Lately, a new static resistivity image reconstruction algorithm is proposed utilizing internal current density data obtained by Magnetic Resonance Current Density Imaging (MRCDI) technique. This new imaging method is called Magnetic Resonance Electrical Impedance Tomography (MREIT). In this paper, we present experimental procedures, denoising techniques, and image reconstructions using a 0.3 Tesla experimental MREIT system and saline phantoms. MREIT using J-substitution algorithm effectively utilizes the internal current density information resolving the problem inherent in a conventional EIT, that is, the low sensitivity of boundary measurements to any changes of internal tissue resistivity values.

Keywords

  • Internal current density
  • J-substitution algorithm
  • MREIT
  • Resistivity image

Fingerprint

Dive into the research topics of 'Magnetic Resonance Electrical Impedance Tomography (MREIT): Phantom experiments for static resistivity images using J-substitution algorithm'. Together they form a unique fingerprint.